Table 1.
Configuration state details (shown on display panel, default in BOLD)
Config State
Description
Options
PAT 2E31
PRBS generator pattern length
2E31
,
2E23
,
2E15
,
2E10
,
2E7
MS
0.500
PRBS generator mark space density
0.500
(1/2)
,
0.250
(1/4),
0.125
(1/8)
Jitter
0
Jitter injection mode
0
(low-jitter mode)
,
1
(external jitter
on)
Synth
1
Internal clock mode
0
(internal clock off),
1
(Internal clock
on)
Err
00000
Lowest digits of error counter
ErAd
OFF
PRBS digital error injection
OFF
,
1E0
,
1E1
, …,
1E7
(errors per
second)
NoData 0
Error detector data sense
0
(data sense off)
,
1
(data sense on)
TD= 0010
Time delay for measurement start (for
long length DUT)
10
-9999 (1/10
th
of a microsecond)
APT=0010
Autophase procedure measurement
time
0-5000 (1/10
th
of a millisecond per
phase step; default
10
(1ms time per
step))
1234R567
Serial number and FW revision (only
the last four digits of the serial number
are shown on the front panel display)
none
(Display > Scroll to select state; Adjust > Config State to change value)
Содержание N4962A
Страница 1: ...Agilent N4962A Serial BERT 12 5 Gb s User Guide...
Страница 6: ......
Страница 16: ...Getting Started 16 Serial BERT 12 5 Gb s User Guide...
Страница 24: ...N4962A System Overview 24 Serial BERT 12 5 Gb s User Guide...
Страница 36: ...System Details and Performance Specifications 36 Serial BERT 12 5 Gb s User Guide...
Страница 60: ...Operation 60 Serial BERT 12 5 Gb s User Guide...
Страница 86: ...Remote GPIB Interface 86 Serial BERT 12 5 Gb s User Guide...
Страница 88: ...Copyright Agilent Technologies 2012 Third edition May 2013 Printed in Germany...