![Agilent Technologies N4962A Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/agilent-technologies/n4962a/n4962a_user-manual_2868170051.webp)
Operation
Serial BERT 12.5 Gb/s User Guide
51
4.3
Rear Panel Interface
Figure 13. N4962A rear panel
The N4962A rear panel features connectors for the internal and external clock
system, as well as a PRBS pattern trigger, and a differential pair of connectors
for a unique data path (DataThru). The rear panel also includes a 5-bit GPIB
address switch, and a GPIB connector. The panel also features the N4962A
power switch, and a built-in cooling fan.
The DataThru/DataThru
¯¯¯¯¯¯ connectors (both SMA) are connected to the IN/IN
¯¯
connectors when the Data Path option is set to Thru (detailed in Section 2.5).
The PRBS generator is enabled and is available from the OUT/OUT
¯¯¯ connectors
when this data path is selected. The input signal is not processed by the error
detector, and errors are not counted.
The Pat TrigO connector is an output pattern trigger from the PRBS generator.
The Pat TrigO signal changes state after 32 complete PRBS bit streams have
been generated. Pat TrigO has a frequency equal to (clock_rate) / (32 *
bit_pattern_length); With a clock speed of 10 GHz, the Pat TrigO frequency is
2.4 MHz with a pattern length of 2
7
-1, and 0.15 Hz with a pattern length of
2
31
-1.
TX CKI and RX CKI are the transmit (PRBS generator) and receiver (error
detector) clock inputs. To use an external 500 Mb/s to 12.5 Gb/s clock signal,
connect the external clock to both TX CKI and RX CKI, as described in Sections
3.4 and 3.5, respectively.
Содержание N4962A
Страница 1: ...Agilent N4962A Serial BERT 12 5 Gb s User Guide...
Страница 6: ......
Страница 16: ...Getting Started 16 Serial BERT 12 5 Gb s User Guide...
Страница 24: ...N4962A System Overview 24 Serial BERT 12 5 Gb s User Guide...
Страница 36: ...System Details and Performance Specifications 36 Serial BERT 12 5 Gb s User Guide...
Страница 60: ...Operation 60 Serial BERT 12 5 Gb s User Guide...
Страница 86: ...Remote GPIB Interface 86 Serial BERT 12 5 Gb s User Guide...
Страница 88: ...Copyright Agilent Technologies 2012 Third edition May 2013 Printed in Germany...