7-12
Agilent B2200 User’s Guide, Edition 2
Error Messages
B2200 Specific Error Messages
B2200 Card/Mode/Port Related Errors
Error
Number
Error Message and Description
3000
Card0 initialization fail
The B2200 may be defective. Contact your nearest Agilent
Technologies service center.
3001
Card1 initialization fail
The B2200 or card installed in slot 1 of the B2200 may be defective.
3002
Card2 initialization fail
The B2200 or card installed in slot 2 of the B2200 may be defective.
3003
Card3 initialization fail
The B2200 or card installed in slot 3 of the B2200 may be defective.
3004
Card4 initialization fail
The B2200 or card installed in slot 4 of the B2200 may be defective.
3011
Bad couple port number
For the
Input Port No.
for the couple port, only 1, 3, 5, 7, or 9 are
allowed.
3012
Bad bias port number
For
Input Port No.
for Bias Port, only 1 to 10 are allowed.
3013
Cannot connect multiple channels in SROUTe mode
For the single connection rule, an input port can be connected to only
one output port, and an output port can be connected to only one
input port.
3014
Cannot directly specify Bias Port channel
Relays on the input bias port cannot be controlled directly when the
bias mode is ON.
Содержание B2200A
Страница 17: ...1 Introduction ...
Страница 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Страница 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Страница 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Страница 35: ...2 Installation ...
Страница 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Страница 61: ...3 Front Panel Operation ...
Страница 99: ...4 Programming ...
Страница 127: ...5 SCPI Command Reference ...
Страница 187: ...6 VXI plug play Driver ...
Страница 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Страница 221: ...7 Error Messages ...