Agilent B2200 User’s Guide, Edition 2
3-11
Front Panel Operation
Operation
To Set Remote Display Mode
This instruction enables or disables the data display in the GPIB remote condition.
See “RMT_DSPL” on page 3-33.
LCD
1. Press the
Menu
key. The setup menu will be displayed on the LCD.
2. Move the cursor to RMT_DSPL, then press the
Enter
key.
3. Move the cursor to LCD, then press the
Enter
key.
4. Press the arrow keys to set the remote display mode ON or OFF.
5. Press the
Enter
key to make the setup effective, or press the
Exit
key to cancel
changing the setup.
LED matrix
1. Press the
Menu
key. The setup menu will be displayed on the LCD.
2. Move the cursor to RMT_DSPL, then press the
Enter
key.
3. Move the cursor to LED, then press the
Enter
key.
4. Press the arrow keys to set the remote display mode ON or OFF.
5. Press the
Enter
key to make the setup effective, or press the
Exit
key to cancel
changing the setup.
To Return to Local Mode
Press the
Local
key. If the front panel keys are locked, send the :SYST:KLC
command from an external computer, then press the
Local
key.
Содержание B2200A
Страница 17: ...1 Introduction ...
Страница 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Страница 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Страница 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Страница 35: ...2 Installation ...
Страница 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Страница 61: ...3 Front Panel Operation ...
Страница 99: ...4 Programming ...
Страница 127: ...5 SCPI Command Reference ...
Страница 187: ...6 VXI plug play Driver ...
Страница 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Страница 221: ...7 Error Messages ...