Agilent B2200 User’s Guide, Edition 2
5-9
SCPI Command Reference
*RST
*RST
This command performs an instrument reset. Status after *RST is shown below:
Channel Configuration:
Auto configuration mode
Connection Rule:
Free
Connection Sequence:
Break Before Make
Bias Mode:
Off
Bias Input Port:
10
Bias-enabled Channels:
All channels (output ports) are bias-enabled.
Couple Mode:
Off
Couple Input Port:
Cleared.
Ground Mode:
Off
Ground Input Port:
12
Ground-enabled Channels:
Cleared.
Ground-enabled Ports:
Cleared.
Card Channel Status:
All relays are opened.
Self-Test Result:
Not changed.
Input Port Symbol String:
Cleared.
Output Port Symbol String:
Cleared.
Beeper:
On
LCD in remote mode:
Off
LED in remote mode:
On
Light Pen:
Enabled.
Syntax
*RST
Example
OUTPUT @Agb2200;"*RST"
NOTE
This command does not change the self-test result.
For the power-on settings, see :SYSTem:CPON.
Содержание B2200A
Страница 17: ...1 Introduction ...
Страница 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Страница 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Страница 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Страница 35: ...2 Installation ...
Страница 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Страница 61: ...3 Front Panel Operation ...
Страница 99: ...4 Programming ...
Страница 127: ...5 SCPI Command Reference ...
Страница 187: ...6 VXI plug play Driver ...
Страница 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Страница 221: ...7 Error Messages ...