4-10
Agilent B2200 User’s Guide, Edition 2
Programming
Programming Examples
Using Bias Mode
The following example uses the bias mode.
Figure 4-3
Bias Mode Example
Setup:
•
Channel configuration mode: Auto
•
Connection rule: Single. Multiple connection is available for the bias port.
•
Connection sequence: Break_Before_Make
•
Used module: All switch modules installed in the mainframe.
•
Bias port: AUX Input 10
•
Bias enabled output port: All output ports
•
Connection paths (connected when the bias mode is ON):
from bias port to all bias enabled output ports
•
Connection paths (connected by the :ROUT:CLOS command):
The following output ports will be disconnected from the bias port, and the
following input-output connections will be made.
from SMU1 to Output 1 (channel list 00101)
from SMU2 to Output 2 (channel list 00202)
from SMU3 to Output 3 (channel list 00303)
from SMU4 to Output 4 (channel list 00404)
Содержание B2200A
Страница 17: ...1 Introduction ...
Страница 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Страница 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Страница 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Страница 35: ...2 Installation ...
Страница 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Страница 61: ...3 Front Panel Operation ...
Страница 99: ...4 Programming ...
Страница 127: ...5 SCPI Command Reference ...
Страница 187: ...6 VXI plug play Driver ...
Страница 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Страница 221: ...7 Error Messages ...