:SYSTem:MEMOry:COMMent
:SYSTem:MEMOry:COMMent
This command memorizes the comment for the B2200 setup information specified
by
memory_number
. The previous comment will be deleted. The query returns the
comment for the specified setup.
Syntax
:SYSTem:MEMOry:COMMent
memory_number
,'
comment
'
:SYSTem:MEMOry:COMMent?
memory_number
Example
OUTPUT @Agb2200;":SYST:MEMO:COMM 1,'1-1,2-13,3-25'"
OUTPUT @Agb2200;":SYST:MEMO:COMM? 1"
ENTER @Agb2200;A$
In this example,
A$
will be
1-1,2-13,3-25
.
:SYSTem:MEMOry:DELete
This command deletes the B2200 setup information and the comment specified by
memory_number
.
Syntax
:SYSTem:MEMOry:DELete
memory_number
Example
OUTPUT @Agb2200;":SYST:MEMO:DEL 1"
Parameter Explanation
memory_number
memory number: 1 to 8.
comment
Comment. Maximum 16 characters. If this command is
entered with empty
comment
, the comment will be cleared.
Numeric characters, alphabetic characters; upper case and
lower case, space, and the following symbols are available.
! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { | } ^ _
Parameter Explanation
memory_number
memory number: 1 to 8.
Содержание B2200A
Страница 17: ...1 Introduction ...
Страница 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Страница 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Страница 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Страница 35: ...2 Installation ...
Страница 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Страница 61: ...3 Front Panel Operation ...
Страница 99: ...4 Programming ...
Страница 127: ...5 SCPI Command Reference ...
Страница 187: ...6 VXI plug play Driver ...
Страница 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Страница 221: ...7 Error Messages ...