Agilent B2200 User’s Guide, Edition 2
6-13
VXI plug&play Driver
Driver Functions
agb220xa_cmdData_Q
This function passes the cmd_str string to the instrument. This entry point will wait
for a response which may be any data. You specify the cmd_str and size parameters,
and get result[ ].
Syntax
ViStatus _VI_FUNC agb220xa_cmdData_Q(ViSession vi, ViString cmd_str,
ViInt32 size, ViChar_VI_FAR result[ ] );
Parameters
vi
Instrument handle returned from agb220xa_init( ).
cmd_str
Instrument command (cannot exceed 256 bytes in length).
size
Length of result in bytes. 2 to 32767.
result[ ]
Response from instrument.
agb220xa_cmdInt
This function passes the cmd_str string to the instrument. This entry point passes the
string in cmd_str followed by a space and then the integer in value. Note that either
an Int16 or 32 can be passed as the Int16 will be promoted.
Syntax
ViStatus _VI_FUNC agb220xa_cmdInt(ViSession vi, ViString cmd_str,
ViInt32 value);
Parameters
vi
Instrument handle returned from agb220xa_init( ).
cmd_str
Instrument command (cannot exceed 256 bytes in length).
value
Parameter for command. -2147483647 to 2147483647.
Содержание B2200A
Страница 17: ...1 Introduction ...
Страница 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Страница 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Страница 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Страница 35: ...2 Installation ...
Страница 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Страница 61: ...3 Front Panel Operation ...
Страница 99: ...4 Programming ...
Страница 127: ...5 SCPI Command Reference ...
Страница 187: ...6 VXI plug play Driver ...
Страница 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Страница 221: ...7 Error Messages ...