Agilent B2200 User’s Guide, Edition 2
5-23
SCPI Command Reference
:DIAGnostic:TEST:FRAMe[:EXECute]?
:DIAGnostic:TEST:FRAMe[:EXECute]?
This command executes the specified B2200 test, then returns the test result. See
“Selftest Menu” on page 3-34 to perform the test.
After the controller test, the B2200 status becomes same as after *RST command
execution.
Syntax
:DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FPANel |
LED | PEN | BEEPer
Query response
test_result
<newline><^END>
1
: fail
0
: pass
Example
OUTPUT @Agb2200;":DIAG:TEST:FRAM? CONT"
Parameter Explanation
CONTroller
Controller test
FPANel
Front panel interface test
LED
LED matrix test
PEN
Light pen test
BEEPer
Beeper test