5-36
Agilent B2200 User’s Guide, Edition 2
SCPI Command Reference
[:ROUTe]:CONNection:SEQuence
[:ROUTe]:CONNection:SEQuence
This command specifies the connection sequence mode, which is the open/close
sequence of the relays when connection route is changed from an existing
connection to a new connection. The query returns the connection sequence mode of
the specified card. At *RST, BBMake is selected. See “Connection Sequence” on
page 3-15.
The connection sequence is used only for cards that have been set to SROUte
connection rule. Also see “[:ROUTe]:CONNection:RULE”.
Syntax
[:ROUTe]:CONNection:SEQuence
card_number
,
sequence
[:ROUTe]:CONNection:SEQuence?
card_number
Query response
NSEQ
or
BBM
or
MBBR
<newline><^END>
Example
OUTPUT @Agb2200;":ROUT:CONN:SEQ ALL,MBBR"
OUTPUT @Agb2200;":ROUT:CONN:SEQ? 0"
ENTER @Agb2200;A$
In this example,
A$
will be
MBBR
.
Parameter Explanation
card_number
For Auto Config mode:
0
or
ALL
For Normal Config mode:
1
,
2
,
3
,
4
, or
ALL
ALL
is not available for query.
sequence
NSEQ
: No-Sequence mode
BBM
: Break-Before-Make mode
MBBR
: Make-Before-Break mode
Содержание B2200A
Страница 17: ...1 Introduction ...
Страница 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Страница 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Страница 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Страница 35: ...2 Installation ...
Страница 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Страница 61: ...3 Front Panel Operation ...
Страница 99: ...4 Programming ...
Страница 127: ...5 SCPI Command Reference ...
Страница 187: ...6 VXI plug play Driver ...
Страница 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Страница 221: ...7 Error Messages ...