2-24
Agilent B2200 User’s Guide, Edition 2
Installation
Measurement Cable Length
Table 2-8
Guard Capacitances of B2200 Measurement Environment (Typical)
Connection
Agilent Model/Part No.
Cable
Length
Guard
Capacitance
from SMU to B2200
inputs
16494A-003 (for non-Kelvin)
80 cm
75 pF
16494A-001 (for non-Kelvin)
1.5 m
130 pF
16494B-003 (for Kelvin, 4142B)
80 cm
90 pF
16493K-001 (for Kelvin)
1.5 m
150 pF
mainframe and
modules
B2210A (only 1 card is installed)
−
145 pF
B2210A (more 1 card)
−
8 pF/extra card
B2211A (only 1 card is installed)
−
145 pF
B2211A (more 1 card)
−
8 pF/extra card
from B2200 outputs
to connector plate or
probe card interface
16494A-001 (for non-Kelvin)
1.5 m
130 pF
16494A-002 (for non-Kelvin)
3 m
240 pF
16494A-003 (for non-Kelvin)
4 m
340 pF
16494B/C-001 (for Kelvin)
1.5 m
140 pF
16494B/C-002 (for Kelvin)
3 m
260 pF
16494C-003 (for Kelvin)
4 m
340 pF
from connector
plate to DUT
Agilent Part No. 8121-1191
X m
130 pF per m
probe card interface
B2220A
−
70 pF
Содержание B2200A
Страница 17: ...1 Introduction ...
Страница 21: ...Agilent B2200 User s Guide Edition 2 1 5 Introduction Front Panel Figure 1 1 Front Panel View ...
Страница 23: ...Agilent B2200 User s Guide Edition 2 1 7 Introduction Rear Panel Figure 1 2 Rear Panel View ...
Страница 34: ...1 18 Agilent B2200 User s Guide Edition 2 Introduction Accessories and Options ...
Страница 35: ...2 Installation ...
Страница 60: ...2 26 Agilent B2200 User s Guide Edition 2 Installation Maintenance ...
Страница 61: ...3 Front Panel Operation ...
Страница 99: ...4 Programming ...
Страница 127: ...5 SCPI Command Reference ...
Страница 187: ...6 VXI plug play Driver ...
Страница 220: ...6 34 Agilent B2200 User s Guide Edition 2 VXI plug play Driver Driver Functions ...
Страница 221: ...7 Error Messages ...