Troubleshooting and Diagnostics
4
34980A Service Guide
85
Instrument Errors
These errors typically do not indicate a hardware failure. They are related to improper
settings usually in command parameters.
Error
Description
111
112
"Channel list: slot number out of range"
"Channel list: channel number out of range"
113
"Channel list: empty scan list"
114
"Channel list: ABus channels not allowed as endpoint in range"
201
"Memory lost: stored state"
202
"Memory lost: power-on state"
203
"Memory lost: stored readings"
204
"Memory lost: time and date"
221
"Settings conflict: calculate limit state forced off"
222
222
"Settings conflict: module type does not match state"
"Settings conflict: Scan list cleared"
223
"Settings conflict: trig source changed to IMM"
224
"Settings conflict: chan adv source changed to IMM"
225
"Settings conflict: DMM disabled or missing"
226
"Settings conflict: DMM enabled"
227
"Settings conflict: DMM not installed"
251
"Unsupported temperature transducer type"
261
"Not able to execute while scan initiated"
262
"Not able to abort scan"
263
"Not able to execute while DMM is measuring"
264
"Not a scannable channel"
271
"Not able to accept unit names longer than 3 characters"
272
"Not able to accept character in unit name"
281
"Not able to perform on more than one channel"
291
"Not able to recall state: it is empty"
292
"Not able to recall state: DMM enable changed"
深圳德标仪器
135-1095-0799
Содержание Agilent 34980A
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...iii 深圳德标仪器 135 1095 0799 ...
Страница 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Страница 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Страница 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Страница 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Страница 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...