Disassembly and Repair
5
34980A Service Guide
121
Front Panel Removal
1
Loosen the front panel assembly by lifting gently on the four plastic ears and moving
the front panel off the mainframe. Stand the entire assembly on its side to make
removal easier.
a
Pull to unclip the ground connector from the mainframe.
b
Swing the two clips securing the front panel connector ribbon cable to the front
panel circuit board and unplug the connector.