Calibration Procedures
3
34980A Service Guide
61
Optional AC Performance Verification Tests
These tests are
not
intended to be performed with every calibration. They are provided as
an aid for verifying additional instrument specifications. There are no adjustments for
these tests; they are provided for performance verification only.
Configuration:
AC Volts
CONFigure[:VOLTage]:AC
LF 3 HZ:SLOW
[SENSe:]VOLTage:AC:BANDwidth 3
1
Make sure you have read
2
Select the AC Volts function and the 3 Hz filter.
3
Select each range in the order shown below. Provide the input voltage and frequency
indicated. Compare measurement results to the appropriate test limits shown in the
table below. (Be certain to allow for appropriate source settling.)
Input Error
from
Nominal
Voltage
Frequenc
y
Range
24 hour
90 day
1 year
1 V
20 Hz
1 V
± 700 µV
± 900 µV
± 1 mV
1 V
20 kHz
1 V
± 700 µV
± 900 µV
± 1 mV
1 V
100 kHz
1 V
± 6.3 mV
± 6.8 mV
± 6.8 mV
1 V
300 kHz
1 V
± 45 mV
± 45 mV
± 45 mV
10 V
1 kHz
10 V
± 7 mV
± 9 mV
± 10 mV
1 V
1 kHz
10 V
± 3.4 mV
± 4.5 mV
± 4.6 mV
100 mV
1 kHz
10 V
± 13 mV
± 14 mV
± 14 mV
深圳德标仪器
135-1095-0799
Содержание Agilent 34980A
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...iii 深圳德标仪器 135 1095 0799 ...
Страница 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Страница 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Страница 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Страница 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Страница 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...