8
34980A Service Guide
1
Obtaining Service
Self Test Procedures
Power–On Self–Test
Each time the instrument is powered on, a subset of self–tests are performed. These tests
check that the minimum set of logic and output hardware are functioning properly.
Complete Self–Test
To perform a complete self–test send the
*TST?
command.
This command performs a complete self-test of the instrument and all installed plug-in
modules and returns a pass/fail indication. The self-test runs a series of tests and,
depending upon the modules installed, may take up to 2 minutes to complete (be sure to
set an appropriate interface time out). If all tests pass, you can have a high confidence
that the instrument and all installed plug-in modules are operational.
If the self–test is successful, SELF–TEST PASSED is displayed on the front panel.
If the self–test fails, SELF–TEST FAILED is displayed and an error number is shown.
Self–test error numbers and their meaning are shown in the table on
•
On the 34945A Microwave Switch/Attenuator Driver, this command performs a
self-test of the 34945A and all connected 34945EXT remote modules.
•
If you have a 34951A Isolated DAC Module installed, the self-test
will require an additional 15 seconds to complete per DAC module
(a memory test is performed).
If one or more tests fail, return the instrument to Agilent for service.
Self Test Error Numbers
On the remote interface, a self–test failure will generate SCPI error –330 and a
supplemental message indicating one of the test numbers shown in the table on
Calibration Errors
The table on
shows failures that may occur during a calibration.
NOTE
The self-test will abort if any signals are connected to ABus1 via the
rear-panel Analog Bus connector (pins 4, 5, and 9). Be sure to
disconnect any signals from ABus1 prior to running the self-test.
深圳德标仪器
135-1095-0799
Содержание Agilent 34980A
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...iii 深圳德标仪器 135 1095 0799 ...
Страница 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Страница 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Страница 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Страница 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Страница 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...