40
34980A Service Guide
2
Specifications
Understanding the “% of range“ Error
The range error compensates for
inaccuracies that result from the function and range you select. The range error
contributes a constant error, expressed as a percent of range, independent of the input
signal level. The following table shows the range error applied to the DMM’s 24-hour dc
voltage specification.
Total Measurement Error
To compute the total measurement error, add the reading
error and range error. You can then convert the total measurement error to a “percent of
input” error or a “ppm (part-per- million) of input” error as shown below.
Example: Computing Total Measurement Error
Assume that a 5 Vdc signal is input to the DMM on the 10 Vdc range.
To compute the total measurement error using the 90-day accuracy specification of
±(0.0020% of r 0.0005% of range).
Range
Input Level
Range Error
(% of range)
Range Error
voltage
10 Vdc
10 Vdc
10 Vdc
10 Vdc
1 Vdc
0.1 Vdc
0.0004
0.0004
0.0004
40
V
40
V
40
V
% of input error =
Total Measurement Error
Input Signal Level
------------------------------------------------------------- 100
ppm of input error =
Total Measurement Error
Input Signal Level
------------------------------------------------------------- 1 000 000
Reading Error
0.0020% x 5 Vdc = 100
V
=
Range error = 0.0005% 10 Vdc = 50
V
Total Error = 100
V + 50
V = 150
V
= 0.0030% of 5 Vdc
= 30 ppm of 5 Vdc
深圳德标仪器
135-1095-0799
Содержание Agilent 34980A
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...iii 深圳德标仪器 135 1095 0799 ...
Страница 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Страница 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Страница 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Страница 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Страница 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...