72
34980A Service Guide
3
Calibration Procedures
Analog Output Adjustment
Install the 34951A module in the mainframe and allow a 1 hour warm-up before
performing these procedures.
This adjustment procedure sets a zero adjustment and a gain adjustment constant for
each DAC output. You must perform all the adjustments on one analog output channel
before adjusting the other analog output channel.
1
Install the module(s) in the instrument. Remove any inputs from the ABus connector.
2
Set the calibration security for the adjustment mode you desire, see
3
Send the
CALibration:MODule? <slot>
command to begin the auto-calibration
procedure for all channels on the module in
<slot>
. You may use the
CALibration:MODule? ALL
command to perform the auto-calibration on all
channels for all modules installed in the mainframe. The complete procedure may
take up to 1 minute per module. Be sure to set the remote command time-out value
appropriately.
4
The command returns a value of “+0” if the calibration is successful.
A returned value of “+1” indicates a calibration failure.
CAUTION
Remove any ABus connector before performing this procedure.
NOTE
Following the auto-calibration procedure, the DMM is left in its reset
state (DCV).
深圳德标仪器
135-1095-0799
Содержание Agilent 34980A
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...iii 深圳德标仪器 135 1095 0799 ...
Страница 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Страница 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Страница 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Страница 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Страница 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...