34980A Service Guide
3
Calibration Procedures
Analog Output Verification Test
This procedure is used to check the calibration of
the analog outputs on the 34951A 4-channel DAC Module. Verification checks are
performed only for those output values with unique calibration constants.
1
Using the input test connector described on
, leave the current input terminal
open. Set the DMM to the 100 mA range. Make and record a current measurement.
This value will be used as an offset value during the output current zero verification.
2
Make voltage measurement connections to channel 1 of the module.
3
Set each output voltage value in the table below and make a measurement. Compare
measurement results to the appropriate test limits shown in the table.
4
Make current measurement connections to channel 1 of the module. Set the DMM to
the 100 mA range.
5
Set each output current value in the table below and make a measurement. Compare
measurement results to the appropriate test limits shown in the table. When
measuring the 0 mA output value, apply the offset value from step 1 to the measured
value.
6
Repeat steps 2 through 5 for channels 2, 3 and 4.
7
Remove the input test connector.
NOTE
It is not necessary to test the voltage output at the full rated 10 mA load.
If you test the output using a load, connect the sense terminals.
深圳德标仪器
135-1095-0799
Содержание Agilent 34980A
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...iii 深圳德标仪器 135 1095 0799 ...
Страница 16: ...xiii 34980A Service Guide 深圳德标仪器 135 1095 0799 ...
Страница 26: ...10 34980A Service Guide 1 Obtaining Service 深圳德标仪器 135 1095 0799 ...
Страница 60: ...44 34980A Service Guide 2 Specifications 深圳德标仪器 135 1095 0799 ...
Страница 94: ...78 34980A Service Guide 3 Calibration Procedures 深圳德标仪器 135 1095 0799 ...
Страница 150: ...134 34980A Service Guide 6 Replaceable Parts 深圳德标仪器 135 1095 0799 ...