Calibration Procedures
3
34980A Service Guide
71
.
[1] Apply a measured “0” offset to this measurement.
Output Voltage
Quick
Check
Error from
Nominal (90 day)
16 V
Q
± 11 mV
12 V
± 9 mV
10 V
± 8 mV
8 V
± 7 mV
4 V
± 5 mV
0 V
Q
± 3 mV
-4V
± 5 mV
-8 V
± 7 mV
-10 V
± 8 mV
-12 V
± 9 mV
-16 V
Q
± 11 mV
Output Current
Quick
Check
Error from
Nominal (90 day)
20 mA
Q
± 23
A
15 mA
± 18.5
A
10 mA
± 14
A
5 mA
± 9.5
A
0 mA
[1]
Q
± 5
A
-5 mA
± 9.5
A
-10 mA
± 14
A
-15 mA
± 18.5
A
-20 mA
Q
± 23
A