118
34980A Service Guide
5
Disassembly and Repair
Power Supply Removal
1
Remove the two T20 Torx screws securing the power supply module.
2
Swing the module out and disengage it from the sheet metal at
the back.
Note:
For testing purposes, y
ou can stand the
power supply on end and insert it into slots
in the mainframe as shown.