7-8
Specifications and Regulatory Information
Specifications
Agilent 86140B, 86143B,
86144B
Agilent 86142B, 86145B,
86146B
Agilent 86141B,
86140B W/Opt 025,
86143B w/Opt 025
Flatness
k
1290-1330 nm
1525-1570 nm
1525-1610 nm
1250-1610 nm
l
-----
±
0.2 dB
-----
-----
±
0.2 dB
-----
±
0.2 dB
±
0.7 dB
-----
±
0.2 dB
-----
-----
Polarization Dependence
m
,
n
1310 nm
±
0.25 dB
±
0.12 dB
----
1530 nm, 1565 nm
±
0.2 dB
±
0.05 dB
----
1600 nm
±
0.25 dB
±
0.08 dB
----
1250 to 1650 nm
±
0.3 dB
±
0.25 dB
±
0.5 dB
a. The 86144B and 86146B specifications are for the 50
µ
m internal path only.
b. Sensitivity is defined as signal value >6 X RMS noise value
c. Temperature range 0 to 30
°
C
d. Second Order
e. Temperature range 20 to 30
°
C
f. Resolution bandwidth setting <channel spacing
g. Characteristic
h. For resolution
≥
0.1 nm
i. Excluding amplitude errors at low power levels due to noise
j. For signals within 8 dB of top of screen
k. With applied input fiber 9/125
µ
m
l. Absorption of light by atmospheric moisture affects flatness at 1350-1420 nm
m. For resolution
≥
0.2 nm
n. Room temperature
Amplitude
Agilent 8614xB
a
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...