3-36
Function Reference
Log Math F=C–D
Log Math F=C
–
D
Subtracts Trace D from Trace C point by point, then stores the results in Trace
F in logarithmic units. If Trace C is in View On mode, this function is continu-
ous and occurs every sweep. Results of the subtraction are referenced to 0 dB.
This often requires a reference level adjustment. For example, if the reference
level is set at –40 dBm, Trace D is at –50 dBm, and Trace C is at
–55 dBm, the F=C–D function places Trace F at –5 dB. Because this is off
screen, the reference level will need to be adjusted.
Key Path
Traces > Trace Math OFF > Default Math Trace F > Log Math F=C–D
Related Functions
Log Math C=A–B, Log Math C=A+B
Remote
Commands
CALCulate6:MATH:EXPRession(TRC/TRD)
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...