![Agilent Technologies 86140B Series Скачать руководство пользователя страница 280](http://html.mh-extra.com/html/agilent-technologies/86140b-series/86140b-series_user-manual_2867952280.webp)
7-19
Specifications and Regulatory Information
Specifications
Option 006 Wavelength Calibrator
The wavelength calibrator option provides an onboard wavelength reference
that can be used to automatically calibrate the optical spectrum analyzer. The
calibrator is based on an EELED and an Acetylene gas absorption cell.
Wavelength Calibrator Block Diagram
The acetylene absorbs light at very specific wavelengths based on the molecu-
lar properties of gas. The cell is illuminated by an EELED and the OSA uses
the absorption pits to perform a wavelength calibration. Since the absorption
of the acetylene gas is a physical constant, it never needs calibrating.
Wavelength calibrator absorption spectrum
The wavelength calibrator enhances the OSA to achieve better than
±
10 pm
wavelength accuracy and removes the need to use a tunable laser source and
multi-wavelength meter as an external reference.
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...