7-10
Specifications and Regulatory Information
Specifications
Table 7-6. Sweep Specifications
Table 7-7. Pulse Mode Accuracy Specifications
Sweep
Agilent 8614xB
a
a. The 86144B and 86146B specifications are for the 50
µ
m internal path only.
Maximum Sweep Rate
b
b. Characteristic
40 nm/56.3 ms
Maximum Sampling Rate in
Zero Span
50
µ
s/trace point
Sweep Cycle Time
50 nm span, auto zero off
50 nm span, auto zero on
100 nm span
500 nm span
<180 ms
<340 ms
<400 ms
<650 ms
ADC Trigger Accuracy
Jitter (distributed uniformly)
Trigger delay range
<
±
0.5
µ
s
2
µ
s-6.5 ms
PULSE MODE ACCURACY
Agilent 86140B, 86143B,
86144B
a
a. The 86144B and 86146B specifications are for the 50
µ
m internal path only.
Agilent 86142B, 86145B,
86146B
Agilent 86141B and
86140B Option 025,
86143B w/Opt 025
Turn On (
≥
2
µ
s after rising
edge)
b
,
c
b. Characteristic
c. Starting from dark
<
±
0.2 dB
<
±
0.2 dB
<
±
0.2 dB
Turn Off (
≥
10
µ
s after falling
edge)
d
d. Characteristic for
86140B, 86141B, 86143B, 86144B, 86146B, 86140B w/Opt 025, 86143B w/Opt 025
<
±
0.2 dB
<
±
0.2 dB
(30 dB extinction)
<
±
0.2 dB
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...