3-57
Function Reference
Peak to Center
Peak to Center
Finds the highest amplitude trace point and sets the center wavelength to that
wavelength.
Key Path
Wavelength > Peak to Center
Remote
Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:MAXimum
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:SCENter
SENSe:WAVelength:CENTer
Peak Excursion (Marker Setup Panel)
Sets the peak excursion value for the marker search functions.
For marker search functions, a signal peak is defined as a rise and fall in the
displayed response by at least the peak excursion value. If peak excursion is
set too high, legitimate peaks may not be discerned as signals. With lower val-
ues more signals will be discerned, but peaks in the noise floor may also be
discerned as signals. To reduce the noise floor variance to a value less than the
peak excursion rate, try using video averaging and/or lowering the sensitivity.
Key Path
Markers > Marker Setup > Peak Excursion
Related Functions
Marker Search Threshold, Next Peak Down
↓
, Next Peak Right
→
, Next
Peak Left
←
Remote
Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:PEXCursion:PEAK
Peak excursion criteria
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...