3-81
Function Reference
Select Path INT/EXT
Select Path INT/EXT
For 86144B/86146B only.
Allows you to select between internal (50
µ
m) and external (9
µ
m) fiber. The
dual fiber outputs from the monochromator allow the instrument to have
improved resolution bandwidth and dynamic range when using the 9
µ
m fiber.
The 50
µ
m fiber is the default setting.
Internal (50
µ
m) path
resolution bandwidth include: 0.06, 0.1, 0.14, 0.2,
0.33, 0.5, 1, 2, 5, 10 settings
External (9
µ
m) path
resolution bandwidth include: 0.04, 0.06, 0.1, 0.14,
0.2, 0.33, 0.5, 1, 2, 5, 10 settings
Key Path
Bandwidth/Sweep > Select Path
Remote
Commands
INSTrument:SELect “FILTER”
Sensitivity
Requests the lowest amplitude signal that can be measured relative to the
power at “top of screen”. It is defined as the signal that is six times the RMS
noise. The minimum setting is –100 dB. An error will be reported for values
outside of this range and the sensitivity will round to the nearest valid sensitiv-
ity.
See “Setting the Sensitivity” on page 2-9
Manual
allows manual input of sensitivities and enables auto gain ranging.
The “top of screen” and the sensitivity setting determines the requested
dynamic range. The system will sweep once per gain stage and may require up
to three sweeps to achieve the requested dynamic range.
Auto
automatically chooses a sensitivity and a single gain range based on “top
of screen”. This will result in approximately 40 dB of dynamic range.
The sweep time (ST) that is displayed in the lower portion of the display is the
time for the OSA to sweep over one gain stage. The OSA may take up to three
sweeps in three different gain stages to make the measurement. This depends
on the settings for sensitivity, reference level, auto range and also the particu-
lar device being measured. The final data trace is a blended composite of each
trace taken in the different gain stages. For more information
.
Содержание 86140B Series
Страница 2: ...Agilent 86140B Series Optical Spectrum Analyzer User s Guide ...
Страница 11: ...1 4 Getting Started Product Overview Agilent 86140B Front and Rear Panels ...
Страница 12: ...1 5 Getting Started Product Overview ...
Страница 14: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 28: ...1 21 Getting Started The Softkey Panels ...
Страница 33: ...1 26 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 41: ...1 34 Getting Started Product Options and Accessories ...
Страница 79: ...2 38 Using the Instrument Analyzer Operating Modes ...
Страница 80: ...3 Function Reference ...
Страница 239: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 283: ...7 22 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...