Error
No.
Display
ed
Message
Description
!
Solution
0310
System
error
Sev
ere
error.
!
Con
tact
y
our
nearest
Agilent
T
ec
hnologies
oce.
0311
Memory
error
Sev
ere
error.
!
Con
tact
y
our
nearest
Agilent
T
ec
hnologies
oce.
Execution
Error
0211
T
rigger
ignored
The
4284A
T
riggered
b efore
the
previous
trigger
w
as
executed.
!
Widen
the
time
interv
al
b et
w
een
triggers.
0222
Data
out
of
range
Data
is
out
of
the
setting
range.
!
En
ter
a
v
alue
within
the
a
v
ailable
setting
range.
0230
Data
corrupt
or
stale
FETCh?
query
receiv
ed
after
the
ABORt
or
the
*RST
command
w
as
receiv
ed,
or
after
the
p o
w
er
on
reset
w
as
p erformed.
!
Send
the
FETCh?
query
during
idle
state
and
when
the
data
is
v
alid
(after
a
measuremen
t
is
p erformed).
FETCh?
query
receiv
ed
after
INITiate
command
w
as
receiv
ed,
and
a
trigger
w
as
not
receiv
ed.
!
Send
the
FETCh?
command
after
the
trigger
is
receiv
ed.
FETCh?
query
receiv
ed
after
the
trigger
w
as
receiv
ed
in
the
list
sw
eep
mo
de
and
the
table
w
as
not
setup.
!
Setup
the
list
sw
eep
table.
Setting
v
alue
query
(for
example,
VOLTage? ,
BIAS:CURRent? ,
LIST:VOLTage?
)
receiv
ed
when
the
setting
mo
de
is
mismatched.
!
Send
a
setting
v
alue
query
command
whic
h
matc
hes
the
curren
t
setting
mo
de.
B-8
Error
and
W
arning
Messages
Содержание 4284A
Страница 12: ......
Страница 37: ...Figure 1 1 Power Cable Supplied Installation and Set Up Guide 1 3 ...
Страница 54: ...Figure 2 5 Display Pages 1 3 2 12 Overview ...
Страница 55: ...Figure 2 5 Display Pages 2 3 Overview 2 13 ...
Страница 56: ...Figure 2 5 Display Pages 3 3 2 14 Overview ...
Страница 58: ...Figure 2 7 Softkey Selection Example 2 16 Overview ...
Страница 60: ...Figure 3 1 Available Fields on the MEAS DISPLAY Page 3 2 DISPLAY FORMAT Menu ...
Страница 61: ...Figure 3 2 Available Softkeys on the MEAS DISPLAY Page DISPLAY FORMAT Menu 3 3 ...
Страница 66: ...Figure 3 4 Effective Measuring Range Oscillator Level 2V or 20 mA 3 8 DISPLAY FORMAT Menu ...
Страница 91: ...Figure 3 15 Available Fields on the LIST SWEEP DISPLAY Page DISPLAY FORMAT Menu 3 33 ...
Страница 96: ......
Страница 99: ...Figure 4 2 Available Softkeys on the MEAS SETUP Page MEAS SETUP Menu 4 3 ...
Страница 104: ...Figure 4 5 Available Operating Area for the ALC Function 4 8 MEAS SETUP Menu ...
Страница 113: ...4 Press NNNNNNNNNNN YES to reset the 4284A MEAS SETUP Menu 4 17 ...
Страница 115: ...Figure 4 8 Available Fields on the CORRECTION Page MEAS SETUP Menu 4 19 ...
Страница 116: ...Figure 4 9 Available Softkeys on the CORRECTION Page 4 20 MEAS SETUP Menu ...
Страница 131: ...Figure 4 15 Available Fields on the LIMIT TABLE SETUP Page MEAS SETUP Menu 4 35 ...
Страница 145: ...Figure 4 21 Available Fields on the LIST SWEEP SETUP Page MEAS SETUP Menu 4 49 ...
Страница 152: ......
Страница 158: ...Figure 5 4 Available Fields on the SYSTEM CONFIG Page 5 6 Catalog System Configuration ...
Страница 182: ...Figure 6 12 Measurement Contacts for Test Leads Extension 6 14 Measurement Procedure and Examples ...
Страница 228: ...Figure 7 19 Standard Event Status Register 7 28 Remote Control ...
Страница 240: ......
Страница 361: ...Caution The memory card should be removed before packing the 4284A General Information 9 29 ...
Страница 362: ......
Страница 432: ......
Страница 436: ......
Страница 440: ......
Страница 450: ...Figure G 8 Short Circuit Recovery Delay Times 3 G 10 Transient States Caused by Measurement Condition Changes ...
Страница 460: ......