![Agilent Technologies 4284A Скачать руководство пользователя страница 400](http://html.mh-extra.com/html/agilent-technologies/4284a/4284a_operation-manual_2867874400.webp)
Multimeter
Reading
(Signal
Frequency:
960
kHz,
Hi
PW:
OFF)
P
ASS
[
]
F
AIL
[
]
OSC
LEVEL
MUL
TIMETER
READING
MINIMUM
(0.97 2M.R.00.5
mV)
A
CTUAL
MAXIMUM
(1.03 2M.R.+0.5
mV)
5
mV
10
mV
20
mV
50
mV
100
mV
200
mV
250
mV
500
mV
1
V
2
V
M.R.
is
the
Multimeter
Reading
for
the
4284A's
test
signal
lev
el.
Multimeter
Reading
(Signal
Frequency:
1.25
kHz,
Hi
PW:
ON)
Option
001
ONL
Y
P
ASS
[
]
F
AIL
[
]
Lev
el
Monitor
Reading
(Signal
Frequency:
1.25
kHz,
Hi-PW:
ON)
Option
001
ONL
Y
OSC
LEVEL
MUL
TIMETER
READING
MINIMUM
(0.97 2M.R.00.5
mV)
A
CTUAL
MAXIMUM
(1.03 2M.R.+0.5
mV)
(0.97 2
M.R.
00.5
mV)
(1.03 2
M.R.+0.5
mV)
10
mV
(0.97 2M.R.
00.5
mV)
(1.03 2M
.R.+0.5
mV)
1
V
(0.97 2M.R.
05
mV)
(1.03 2M.R
.+5
mV)
2.5
V
(0.97 2M.R.
05
mV)
(1.03 2M.
R.+5
mV)
20
V
10-38
P
erformance
T
ests
Содержание 4284A
Страница 12: ......
Страница 37: ...Figure 1 1 Power Cable Supplied Installation and Set Up Guide 1 3 ...
Страница 54: ...Figure 2 5 Display Pages 1 3 2 12 Overview ...
Страница 55: ...Figure 2 5 Display Pages 2 3 Overview 2 13 ...
Страница 56: ...Figure 2 5 Display Pages 3 3 2 14 Overview ...
Страница 58: ...Figure 2 7 Softkey Selection Example 2 16 Overview ...
Страница 60: ...Figure 3 1 Available Fields on the MEAS DISPLAY Page 3 2 DISPLAY FORMAT Menu ...
Страница 61: ...Figure 3 2 Available Softkeys on the MEAS DISPLAY Page DISPLAY FORMAT Menu 3 3 ...
Страница 66: ...Figure 3 4 Effective Measuring Range Oscillator Level 2V or 20 mA 3 8 DISPLAY FORMAT Menu ...
Страница 91: ...Figure 3 15 Available Fields on the LIST SWEEP DISPLAY Page DISPLAY FORMAT Menu 3 33 ...
Страница 96: ......
Страница 99: ...Figure 4 2 Available Softkeys on the MEAS SETUP Page MEAS SETUP Menu 4 3 ...
Страница 104: ...Figure 4 5 Available Operating Area for the ALC Function 4 8 MEAS SETUP Menu ...
Страница 113: ...4 Press NNNNNNNNNNN YES to reset the 4284A MEAS SETUP Menu 4 17 ...
Страница 115: ...Figure 4 8 Available Fields on the CORRECTION Page MEAS SETUP Menu 4 19 ...
Страница 116: ...Figure 4 9 Available Softkeys on the CORRECTION Page 4 20 MEAS SETUP Menu ...
Страница 131: ...Figure 4 15 Available Fields on the LIMIT TABLE SETUP Page MEAS SETUP Menu 4 35 ...
Страница 145: ...Figure 4 21 Available Fields on the LIST SWEEP SETUP Page MEAS SETUP Menu 4 49 ...
Страница 152: ......
Страница 158: ...Figure 5 4 Available Fields on the SYSTEM CONFIG Page 5 6 Catalog System Configuration ...
Страница 182: ...Figure 6 12 Measurement Contacts for Test Leads Extension 6 14 Measurement Procedure and Examples ...
Страница 228: ...Figure 7 19 Standard Event Status Register 7 28 Remote Control ...
Страница 240: ......
Страница 361: ...Caution The memory card should be removed before packing the 4284A General Information 9 29 ...
Страница 362: ......
Страница 432: ......
Страница 436: ......
Страница 440: ......
Страница 450: ...Figure G 8 Short Circuit Recovery Delay Times 3 G 10 Transient States Caused by Measurement Condition Changes ...
Страница 460: ......