Handler
Interface
Function
T
est
(Option
202
only)
P
erform
this
test
only
when
troublesho
oting
the
Option
202
handler
in
terface
b oard.
This
test
v
eries
the
Option
202
handler
in
terface
functions.
When
this
test
is
p erformed
the
follo
wing
LEDs
WILL
NOT
turn
ON
b ecause
the
signals
they
represen
t
are
not
used
b
y
the
Option
202
handler
in
terface
b oard.
PHI-,
PLO-,
SREJ-,
UNBAL-
and
ALARM-
Equipment:
Handler
Sim
ulator
Agilent
PN
04278-65001
Cable
Agilent
PN
04278-61635
Procedure:
1.
P
erform
steps
1
through
5
describ ed
\Pro
cedure:".
2.
Disconnect
the
at
cable
from
the
handler
in
terface
b oard.
The
handler
in
terface
b oard
has
bro
wn
and
an
orange
extractors
and
its
lo
cation
is
sho
wn
in
Figure
10-12.
3.
Remo
v
e
the
handler
in
terface
b oard.
Caution
The
in
terface
b oard
con
tains
electronic
comp onen
ts
that
can
b e
damaged
b
y
static
electricit y
through
electrostatic
disc
harge(ESD).
T
o
prev
en
t
ESD
damage,main tain
frequen
t
con
tact
with
an
y
bare
sheet
metal
surface
on
the
c
hassis.
A
grounding
wrist
strap
(or
similar
device)
is
useful
for
this
purp ose.
Handle
the
b oard
carefully
at
all
times.
Av
oid
touc
hing
electronic
comp onen
ts
or
circuit
paths.
Note
Before
p erforming
step
4,
note
the
jump er
settings
in
order
to
return
them
to
the
same
setting
at
the
end
of
this
function
test.
4.
Set
the
jump ers
on
the
handler
in
terface
b oard
the
same
settings
as
when
the
b oard
is
shipp ed
from
the
factory
referring
to
Figure
10-16.
10-30
P
erformance
T
ests
Содержание 4284A
Страница 12: ......
Страница 37: ...Figure 1 1 Power Cable Supplied Installation and Set Up Guide 1 3 ...
Страница 54: ...Figure 2 5 Display Pages 1 3 2 12 Overview ...
Страница 55: ...Figure 2 5 Display Pages 2 3 Overview 2 13 ...
Страница 56: ...Figure 2 5 Display Pages 3 3 2 14 Overview ...
Страница 58: ...Figure 2 7 Softkey Selection Example 2 16 Overview ...
Страница 60: ...Figure 3 1 Available Fields on the MEAS DISPLAY Page 3 2 DISPLAY FORMAT Menu ...
Страница 61: ...Figure 3 2 Available Softkeys on the MEAS DISPLAY Page DISPLAY FORMAT Menu 3 3 ...
Страница 66: ...Figure 3 4 Effective Measuring Range Oscillator Level 2V or 20 mA 3 8 DISPLAY FORMAT Menu ...
Страница 91: ...Figure 3 15 Available Fields on the LIST SWEEP DISPLAY Page DISPLAY FORMAT Menu 3 33 ...
Страница 96: ......
Страница 99: ...Figure 4 2 Available Softkeys on the MEAS SETUP Page MEAS SETUP Menu 4 3 ...
Страница 104: ...Figure 4 5 Available Operating Area for the ALC Function 4 8 MEAS SETUP Menu ...
Страница 113: ...4 Press NNNNNNNNNNN YES to reset the 4284A MEAS SETUP Menu 4 17 ...
Страница 115: ...Figure 4 8 Available Fields on the CORRECTION Page MEAS SETUP Menu 4 19 ...
Страница 116: ...Figure 4 9 Available Softkeys on the CORRECTION Page 4 20 MEAS SETUP Menu ...
Страница 131: ...Figure 4 15 Available Fields on the LIMIT TABLE SETUP Page MEAS SETUP Menu 4 35 ...
Страница 145: ...Figure 4 21 Available Fields on the LIST SWEEP SETUP Page MEAS SETUP Menu 4 49 ...
Страница 152: ......
Страница 158: ...Figure 5 4 Available Fields on the SYSTEM CONFIG Page 5 6 Catalog System Configuration ...
Страница 182: ...Figure 6 12 Measurement Contacts for Test Leads Extension 6 14 Measurement Procedure and Examples ...
Страница 228: ...Figure 7 19 Standard Event Status Register 7 28 Remote Control ...
Страница 240: ......
Страница 361: ...Caution The memory card should be removed before packing the 4284A General Information 9 29 ...
Страница 362: ......
Страница 432: ......
Страница 436: ......
Страница 440: ......
Страница 450: ...Figure G 8 Short Circuit Recovery Delay Times 3 G 10 Transient States Caused by Measurement Condition Changes ...
Страница 460: ......