![Agilent Technologies 4284A Скачать руководство пользователя страница 264](http://html.mh-extra.com/html/agilent-technologies/4284a/4284a_operation-manual_2867874264.webp)
FUNCtion
Subsystem
:IMP
edance[:TYPE]
The
:IMPedance
command
sets
the
measuremen
t
function.
The
:IMPedance?
query
returns
the
curren
t
measuremen
t
function.
Command
Syntax
FUNCtion:IMPedance
2
:TYPE
3
<function >
Where,
<function >
is:
CPD
Sets
function
to
C
p
-D
LPRP
Sets
function
to
L
p
-R
p
CPQ
Sets
function
to
C
p
-Q
LSD
Sets
function
to
L
s
-D
CPG
Sets
function
to
C
p
-G
LSQ
Sets
function
to
L
s
-Q
CPRP
Sets
function
to
C
p
-R
p
LSRS
Sets
function
to
L
s
-R
s
CSD
Sets
function
to
C
s
-D
RX
Sets
function
to
R-X
CSQ
Sets
function
to
C
s
-Q
ZTD
Sets
function
to
Z-
(deg)
CSRS
Sets
function
to
C
s
-R
s
ZTR
Sets
function
to
Z-
(rad)
LPQ
Sets
function
to
L
p
-Q
GB
Sets
function
to
G-B
LPD
Sets
function
to
L
p
-D
YTD
Sets
function
to
Y-
(deg)
LPG
Sets
function
to
L
p
-G
YTR
Sets
function
to
Y-
(rad)
Example
OUTPUT
717;"FUNC:IMP
GB"
!
Set
to
the
G-B
function
Query
Syntax
FUNCtion:IMPedance
2
:TYPE
3
?
Query
Response
Returned
format
is
:
<function ><NL^END>
Example
10
OUTPUT
717;"FUNC:IMP?"
20
ENTER
717;A$
30
A$
40
END
8-24
Command
Reference
Содержание 4284A
Страница 12: ......
Страница 37: ...Figure 1 1 Power Cable Supplied Installation and Set Up Guide 1 3 ...
Страница 54: ...Figure 2 5 Display Pages 1 3 2 12 Overview ...
Страница 55: ...Figure 2 5 Display Pages 2 3 Overview 2 13 ...
Страница 56: ...Figure 2 5 Display Pages 3 3 2 14 Overview ...
Страница 58: ...Figure 2 7 Softkey Selection Example 2 16 Overview ...
Страница 60: ...Figure 3 1 Available Fields on the MEAS DISPLAY Page 3 2 DISPLAY FORMAT Menu ...
Страница 61: ...Figure 3 2 Available Softkeys on the MEAS DISPLAY Page DISPLAY FORMAT Menu 3 3 ...
Страница 66: ...Figure 3 4 Effective Measuring Range Oscillator Level 2V or 20 mA 3 8 DISPLAY FORMAT Menu ...
Страница 91: ...Figure 3 15 Available Fields on the LIST SWEEP DISPLAY Page DISPLAY FORMAT Menu 3 33 ...
Страница 96: ......
Страница 99: ...Figure 4 2 Available Softkeys on the MEAS SETUP Page MEAS SETUP Menu 4 3 ...
Страница 104: ...Figure 4 5 Available Operating Area for the ALC Function 4 8 MEAS SETUP Menu ...
Страница 113: ...4 Press NNNNNNNNNNN YES to reset the 4284A MEAS SETUP Menu 4 17 ...
Страница 115: ...Figure 4 8 Available Fields on the CORRECTION Page MEAS SETUP Menu 4 19 ...
Страница 116: ...Figure 4 9 Available Softkeys on the CORRECTION Page 4 20 MEAS SETUP Menu ...
Страница 131: ...Figure 4 15 Available Fields on the LIMIT TABLE SETUP Page MEAS SETUP Menu 4 35 ...
Страница 145: ...Figure 4 21 Available Fields on the LIST SWEEP SETUP Page MEAS SETUP Menu 4 49 ...
Страница 152: ......
Страница 158: ...Figure 5 4 Available Fields on the SYSTEM CONFIG Page 5 6 Catalog System Configuration ...
Страница 182: ...Figure 6 12 Measurement Contacts for Test Leads Extension 6 14 Measurement Procedure and Examples ...
Страница 228: ...Figure 7 19 Standard Event Status Register 7 28 Remote Control ...
Страница 240: ......
Страница 361: ...Caution The memory card should be removed before packing the 4284A General Information 9 29 ...
Страница 362: ......
Страница 432: ......
Страница 436: ......
Страница 440: ......
Страница 450: ...Figure G 8 Short Circuit Recovery Delay Times 3 G 10 Transient States Caused by Measurement Condition Changes ...
Страница 460: ......