LAN Configuration
13
Initialization Mode Selection
Initialization Mode Selection
The following initialization (init) modes are selectable:
Default IP address for LAN is 192.168.254.11.
DHCP address is the module’s LAN MAC address.
Using Stored
When initialization mode
Using Stored
is selected, the parameters are taken
from the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured
previously by one of the described methods.
Figure 34
Using Stored (Principle)
Table 17
Initialization Mode Switches
SW1
SW2
SW3 SW4
SW5
SW6
Init
Mode
0
0
0
0
0
0
Use Default IP
Address
0
0
0
0
1
0
Use Stored IP Address
0
0
0
1
0
0
Use DHCP
Note: The setting ‘0’ (down) is essential.