176
Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Metering Seal and Piston
15
Reinstall the capillaries.
16
In the
Tools
section of the Agilent Lab Advisor software
system screen exit
Service & Diagnostics >
Maintenance Positions > Change Metering Device
:
click
Next
and wait until the metering device is in Home
position.
17
Select
Back
to finish the procedure.
18
Close the doors.
19
Perform a pressure test.