![Agilent Technologies 1260 Infinity II Скачать руководство пользователя страница 110](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii/1260-infinity-ii_user-manual_2867537110.webp)
110
Agilent 1260 Infinity II Preparative Autosampler User Manual
7
Troubleshooting and Diagnostics
Maintenance Functions
Maintenance Functions
Certain maintenance procedures require the needle arm, metering device, and
gripper assembly to be moved to specific positions to enable easy access to
components. The maintenance functions move these assemblies into the
appropriate maintenance position. In the Data System the ALS maintenance
positions can be selected from the
Maintenance Positions
menu in the
Service &
Diagnostics
display.
In the LabAdvisor Diagnostic SW the ALS maintenance positions can be
selected from the
Service & Diagnostic > Maintenance Positions
menu in the
Diagnosis
display.
User Interface
The functions for the control software are:
Change Needle, Sample Loop or Needle Seat:
The needle moves slowly down and positions the needle arm for easy access to
the needle, the sample loop or the needle seat.
Change Metering Seal or Piston
:
relieves the tension on the metering spring (draws the piston to the outer
position), enabling easy disassembly of the metering head assembly).
Park Arm
:
secures the gripper arm to the park position behind the sampling unit. ready
for transport or shipping of the autosampler.
Change Gripper:
The change gripper function moves the gripper to the front of the autosampler
enabling easy access to the gripper release mechanism.
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...