Agilent 1260 Infinity II Preparative Autosampler User Manual
233
LAN Configuration
13
TCP/IP parameter configuration
TCP/IP parameter configuration
To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
•
IP address
•
Subnet Mask
•
Default Gateway
The TCP/IP parameters can be configured by the following methods:
•
by automatically requesting the parameters from a network-based DHCP
Server (using the so-called Dynamic Host Configuration Protocol). This
mode requires a LAN-onboard Module or a G1369C LAN Interface card, see
•
by manually setting the parameters using Telnet
•
by manually setting the parameters using the Local Controller
The LAN interface differentiates between several initialization modes. The
initialization mode (short form ‘init mode’) defines how to determine the
active TCP/IP parameters after power-on. The parameters may be derived
non-volatile memory or initialized with known default values. The
initialization mode is selected by the configuration switch, see