Agilent 1260 Infinity II Preparative Autosampler User Manual
173
Maintenance
9
Exchange the Metering Seal and Piston
Exchange the Metering Seal and Piston
When
Poor injection-volume reproducibility
Leaking metering device
Tools required
p/n
Description
8710-0510
Wrench open 1/4 — 5/16 inch
8710-2392
4 mm Hex key
8710-2411
Hex key 3 mm12 cm long
01018-23702 Insert
tool
Cleaning tissue and appropriate solvent like isopropanol or methanol
Parts required
p/n
Description
0905-1294
Metering seal, 900 µL
G4267-60462
Piston Assembly, 900 µL
Preparations
Finish any pending acquisition job and in order to avoid leaks, stop the pump running.
1
In the
Tools
section of the Agilent Lab Advisor software
select
Service & Diagnostics > Maintenance Positions
> Change Metering Device
, click
Start
and wait until
the metering device is in maintenance position.
2
Open the front door.