Agilent 1260 Infinity II Preparative Autosampler User Manual
117
Troubleshooting and Diagnostics
7
Maintenance Functions
Automatic Referencing
Automatic Referencing is required to compensate for small deviations in
positioning of the gripper which may occur after disassembling the module for
repair.
The Automatic Referencing procedure requires Calibration Tool
(G7129-60014) and uses several tray positions as reference points. Because
the tray is a rectangle, a four-point alignment is sufficient to correct all other
vial positions within the tray. On completion of the procedure, the corrected
gripper positions are stored in the instrument firmware.
Start
Moves the transport assembly and gripper arm to the position required to
install the calibration tools.
Next
Mount the calibration fixture. Position P1-C6 and P2-F6.
Next
Mount a 2 mL vial.
Next
Perform autoreferencing.
Next
Remove calibration fixture.
Back
Completes the procedure.
N O T E
Please check help for further explanation.
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...