Agilent 1260 Infinity II Preparative Autosampler User Manual
7
Troubleshooting and Diagnostics
Troubleshooting Guide for the Sample Transport Assembly
Poor alignment
With Error messages
• motor overtemp
(0 or 2 or 3)
• movement failed
(0 or 2 or 3)
1
Start automatic referencing in Lab Advisor. For this procedure it is
mandatory to use the calibration tool.
On completion of the procedure, the corrected values are stored in the
instrument firmware.
2
Exchange the gripper arm assembly (Gripper assembly (G1313-60010)), see
3
Exchange Transport assembly ( Sample Transport Assembly
(G7129-60600)).
4
Exchange the Autosampler Main Board ( Autosampler Specific Board
(G7129-65800)) or the fusion board ( Fusion Board (G4200-65880)).
5
For all three steps (step 2, 3, and 4) you have to do an automatic referencing
again to store the correct values in the instrument firmware again.
WA R N I N G
Personal damage, damage to the module
➔
Some of these procedures require a trained service engineer. Persons who are not
qualified MUST NOT perform these procedures.
N O T E
When a motor over temperature message has occurred, the sampler must be turned OFF
for about 10 min to allow the motor to cool down.
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...