![Agilent Technologies 1260 Infinity II Скачать руководство пользователя страница 119](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii/1260-infinity-ii_user-manual_2867537119.webp)
Agilent 1260 Infinity II Preparative Autosampler User Manual
119
Troubleshooting and Diagnostics
7
Step Commands
Step Commands
Each movement of the sampling sequence can be done under manual control.
This is useful during troubleshooting, where close observation of each of the
sampling steps is required to confirm a specific failure mode or verify
successful completion of a repair.
Each injector step command actually consists of a series of individual
commands that move the autosampler components to predefined positions,
enabling the specific step to be done.
Table 10
Injector step commands
Step
Action
Comments
Valve Bypass
Switches injection valve to the bypass
position.
Piston Home
Moves the piston to the home position.
Needle Up
Lifts the needle arm to the upper
position.
Command also switches the
valve to bypass if it is not already
in that position.
Vial to Seat
Moves the selected vial to the seat
position.
Command also lifts the needle to
the upper position.
Needle into Sample
Lowers the needle into the sample.
Command also positions the vial
at the seat, and lifts the needle to
the upper position.
Draw
Metering device draws the defined
injection volume.
Command also positions the vial
at the seat, lifts the needle, and
lowers the needle into vial.
Command can be done more than
once (maximum draw volume of
100 µL cannot be exceeded). Use
Plunger Home
to reset the
metering device.
Needle Up
Lifts the needle out of the vial.
Command also switches the
valve to bypass if it is not already
in that position.
Vial to Tray
Returns the selected vial to the tray
position.
Command also lifts the needle to
the upper position.
Needle into Seat
Lowers the needle arm into the seat.
Command also returns the vial to
the tray position.
Valve Mainpass
Switches the injection valve to the
mainpass position.
Reset
Resets the injector.
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...