Agilent 1260 Infinity II Preparative Autosampler User Manual
12
Hardware Information
Interfaces
5V Distribution (Future Use)
•
Available directly after turn on oft the hosting module (assures that certain
base functionality of the device can be detected by firmware).
•
For digital circuits or similar.
•
Provided 500 mA maximum.
•
Short-circuit proof with automatic switch off (by firmware).
24V Distribution (Future Use)
•
Available by firmware command (defined turn on/off).
•
For devices that need higher power
•
Class 0: 0.5 A maximum (12 W)
•
Class 1: 1.0 A maximum (24 W)
•
Class 2: 2.0 A maximum (48 W)
•
Class depends on hosting module’s internal power overhead.
•
If a connected device requires more power the firmware detects this
(overcurrent detection) and provides the information to the user interface.
•
Fuse used for safety protection (on board).
•
Short circuit will be detected through hardware.
USB (Universal Serial Bus)
USB (Universal Serial Bus) - replaces RS232, supports:
•
a PC with control software (for example Agilent Lab Advisor)
•
USB Flash Disk
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...