Agilent 1260 Infinity II Preparative Autosampler User Manual
12
Hardware Information
Setting the 6-bit Configuration Switch
Table 15
6-bit Configuration Switch
Mode
Function/Setting
Switch 1
Switch 2
Switch 3
Switch 4
Switch 5
Switch 6
COM
1
1
When selecting mode COM, settings are stored to non-volatile memory. When selecting mode TEST, COM settings are
taken from non-volatile memory.
0
n.a.
2
2
not assigned - Always keep these switches on position ‘0’ (off)
n.a.
LAN Init Mode
n.a.
Use Default IP Address
3
3
Default IP Address is 192.168.254.11
0
0
0
0
0
Use Stored IP Address
0
0
0
1
0
Use DHCP to request IP Address
4
4
Host Name will be the MAC address.
0
0
1
0
0
Test
1
System
n.a.
n.a.
n.a.
ColdStart
Boot Main System / Keep Data
0
0
0
0
0
Boot Resident System / Keep Data
1
0
0
0
0
Boot Main System / Revert to
Default Data
0
0
0
0
1
Boot Resident System / Revert to
Default Data
1
0
0
0
1
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...