Agilent 1260 Infinity II Preparative Autosampler User Manual
11
Identifying Cables
Remote Cables
•
5188-8057 ERI to APG and RJ45 (Connector D_Subminiature 15 pin (ERI),
Connector D_Subminiature 9 pin (APG), Connector plug Cat5e (RJ45))
One end of these cables provides a Agilent Technologies APG (Analytical
Products Group) remote connector to be connected to Agilent modules. The
other end depends on the instrument to be connected to.
Table 12
5188-8057 ERI to APG and RJ45
p/n 5188-8057
Pin (ERI)
Signal
Pin (APG)
Active
(TTL)
Pin (RJ45)
10
GND
1
5
1
Start
Request
9
High
2
Stop
8
High
3
Ready
7
High
4
Fraction
Trigger
5
High
4
5
Power on
6
High
6
Shut Down
4
High
7
Start
3
High
8
Prepare
2
High
Ground
Cable
Shielding
NC