1100 Series Nano Pump Service Manual
61
Troubleshooting and Test Functions
2
Index Missing
Index Missing
: Pump channel A
B: Index Missing
: Pump channel B
The encoder index position in the nano pump was not found during
initialization.
During initialization, the first plunger is moved to the mechanical stop. After
reaching the mechanical stop, the plunger reverses direction until the encoder
index position is reached. If the index position is not recognized within a
defined time, the error message is generated.
Probable Causes
•
Disconnected or defective encoder cable.
•
Defective pump drive assembly.
Suggested Actions
✔
Ensure the encoder cable are not damaged or dirty. Make sure the cables
are connected securely to the CSM board.
✔
Exchange the pump drive assembly.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......