74
1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
Ramp 2
The nano pump continues pumping solvent with a flow rate of 150 µl/min.
Channel A delivers for one pump cycle (first, plunger A2 delivers, then plunger
A1), followed by channel B (plunger B2, then plunger B1), both channels with
a stroke of 20 µl. The pressure increase during this phase should be linear.
Large leaks or defects will be characterized by an unstable, non-linear slope.
Ramp 3
Just before the start of the first plateau, plunger A2 delivers with a flow rate
of 50 µl/min for approximately 8 seconds. The system pressure should be 240
bar or higher.
Plateau 1
At plateau 1 plunger A2 delivers with a flow rate of 3 µl/min for 30 seconds.
During this time, the slope should be horizontal or slightly positive (slight
pressure increase). A negative slope indicates a leak rate greater than
3 µl/min.
Ramp 4
Plunger B2 delivers 50 µl/min for approximately 8 seconds.
Plateau 2
Plunger B2 delivers with a flow rate of 3 µl/min for 30 seconds. The slope
should be horizontal or slightly positive. A negative slope indicates a leak rate
greater than 3 µl/min.
Ramp 5
Plunger A1 delivers 50 µl/min for approximately 8 seconds.
Plateau 3
Plunger A1 with a flow rate of 3 µl/min for 30 seconds. The slope should be
horizontal or slightly positive. A negative slope indicates a leak rate greater
than 3 µl/min.
Ramp 6
Plunger B1 delivers 50 µl/min for approximately 7 seconds.
Plateau 4
Plunger B1 delivers with a flow rate of 3 µl/min for approximately 30 seconds.
The slope should be horizontal or slightly positive. A negative slope indicates a
leak rate greater than 3 µl/min. At the end of the fourth plateau, the test is
finished and the nano pump switches off.
Running the Leak Test
Tools required
Wrench 1/4 inch
Parts and materials
required
Restriction Capillary,G1313-87305
Blank nut, 01080-83202
Isopropanol, 500ml
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......