1100 Series Nano Pump Service Manual
6
Control Module Screens for the Nano Pump
Screens available from the Records screen
Records screen
Use the
Esc
-key until you receive
Views
on the
F5
-key. Choose
System
from
the pull-down menu. Use the
F4
-key (
Records
) and select the pump. Errors
are reported either into the
F2
(
System Log
) or
F3
(
Error Log
).
EMF
Early Mainte-
nance Feedback)
Use the
F1
-key (
EMF
) to set EMF
Limits
. Choose menu item
Setup Limits
to
set the amount of pumped solvent or seal wear, when you want to receive a
warning for exceeded limits.
Press
F7
(
Reset
) to reset the counters after exchanging parts that are subject
to wear. Changes must be activated with
F6
(
Done
).
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......