52
1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
Inlet-Valve Missing
Inlet-Valve Missing:
Pump channel A
B: Inlet-Valve Missing:
Pump channel B
The active-inlet valve in the nano pump is missing or defective.
The processor checks the presence of the active-inlet valve connector every
2 seconds. If the connector is not detected by the processor, the error message
is generated.
Probable Causes
•
Disconnected or defective cable.
•
Disconnected or defective connection cable (front panel to CSPM board).
•
Defective active inlet valve.
Suggested Actions
✔
Ensure the pins of the active inlet valve connector are not damaged. Ensure
the connector is seated securely.
✔
Ensure the connection cable is seated correctly. Exchange the cable if
defective.
✔
Exchange the active inlet valve.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......