1100 Series Nano Pump Service Manual
41
Troubleshooting and Test Functions
2
Pressure Below Lower Limit
The system pressure has fallen below the lower pressure limit.
Probable Causes
•
Lower pressure limit set too high.
•
Air bubbles in the mobile phase.
•
Leak.
•
Defective damper.
•
Defective CSM board.
Suggested Actions
✔
Ensure the lower pressure limit is set to a value suitable for the analysis.
✔
Ensure solvents are degassed. Purge the nano pump.
✔
Ensure solvent inlet filters are not blocked.
✔
Inspect the pump head, capillaries and fittings for signs of a leak.
✔
Purge the nano pump. Run a pressure test to determine whether the seals
or other pump components are defective.
✔
Exchange the damper.
✔
Exchange the CSM board.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......