70
1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
The slope and plateau is evaluated automatically. “Evaluating the Results” on
page 71 describes the evaluation and interpretation of the pressure test
results
Running the test from the Control Module
1
Place a bottle of LC-grade isopropyl alcohol in the solvent cabinet and
connect it to channel A2.
2
Block the injection device (port 6) with a blank nut (01080-83202),
3
Set flow for channel A2 to 2500 µl/min and flush the degasser channel for
about 5 minutes.
4
Set flow to 0 µl/min.
5
Connect the signal cable to the analog output at the rear of the nano pump
(only if an integrator is used).
6
Press Execute to initialize the pressure test.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......