1100 Series Nano Pump Service Manual
85
Troubleshooting and Test Functions
2
Evaluating the results
The test results are evaluated automatically.Two pressure plateaus are
checked; one at 20 bar and the other at 300 bar. The appropriate current is
measured. A further criterion is the difference between the two plateaus. The
limits are as follows:
•
< 85% EMPV current for the 20 bar plateau.
•
> 15% EMPV current for the 300 bar plateau.
•
> 15% difference between both EMPV current values.