122
1100 Series Nano Pump Service Manual
3
Repairing the Pump
Exchanging the Pump Seals and Seal Wear-in Procedure
When required:
•
Seals leaking, if indicated by the results of
the leak test (Check both pump heads
individually!)
Tools required:
•
3-mm hexagonal key
•
4-mm hexagonal key
•
1/4 inch wrench
Parts required:
•
Seals (pack of 2) 5063-6589 (standard) or
0905-1420 (for normal phase application)
For the seal wear-in procedure:
•
Restriction capillary (5022-2159)
1 Disassemble the pump head assembly of the leaky
pump head (see “Removing and Disassembling the
Pump Head Assembly” on page 120).
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......