1100 Series Nano Pump Service Manual
129
Repairing the Pump
3
Exchanging the Optional Interface Board
1
Switch off the nano pump at the main power switch. Unplug the pump from
main power.
2
Disconnect cables from the interface board connectors.
3
Loosen the screws. Slide out the interface board from the nano pump.
4
Install the new interface board. Secure screws.
5
Reconnect the cables to the board connector
.
C A U T I O N
The interface board is sensitive to electrostatic discharge. Always use the ESD kit
when handling electronic boards.
When required
Board defective
Part required
BCD (Interface) board, see
Figure 21
Exchanging the Interface Board
BCD (Interface)
board
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......